"Communication Failure" to Agilent instrumentation within Empower 3 software (Guided Troubleshooting) - WKB233387
Article number: 233387
SYMPTOMS
- Cannot communicate within Empower 3
- System shows as Offline within Run Samples
- Cannot run samples
ENVIRONMENT
- Empower 3
- Agilent LC
- Agilent GC
- Agilent ICF
STRATEGY
- Recent PM to system. WKB116101
- Hand held controller in use. WKB243476
- Confirm hardware is not in Resident mode. WKB89646
- Recent hardware configuration change?
- Confirm iP addresses are correct. WKB10818
- Configure iP within DHCP. WKB10193
- Ensure DIP switches are correct. WKB6565
- If Avocent hub is in use, see WKB7425.
- Confirm Network Card (NIC) is not the cause. WKB81695
- If connections use bridge, see WKB30316.
- Is there a change to the configuration? WKB85498
- Ensure that Plate Type is correct in Run Samples. WKB18149
- Does Preconfiguration tool fail? WKB87276
- Delete INSTSRV.DAT file. WKB102291
- Reconfigure system. WKB102235
- Ensure that the signal source is correct in the instrument method. WKB226654
- Ensure that .NET Framework is correct. WKB121784
- Check the Ethernet cables. WKB7997
ADDITIONAL INFORMATION
Refer to specific Release Notes for Agilent Instrument Control Framework Support Layer on waters.com for configuration guidance.
What is the current ICF version to control Agilent LC/GC in Empower ? - WKB1566