Cannot communicate with Agilent 1200 in Empower - WKB10818
Article number: 10818
SYMPTOMS
- Agilent 1200 is "OK? No" in node properties
- IP address of the Agilent 1200 not listed in DHCP
- IP address configured on the Agilent is 192.168.254.11
- IP address configured on the NIC Card is 192.168.0.1
- Dip switches correct per WKB6565
ENVIRONMENT
- Empower
- Agilent 1200
CAUSE
IP address configured incorrectly on the Agilent LC.
FIX or WORKAROUND
- Set the static IP address through the Gameboy for the Agilent 1200 to 192.168.0.2.
- Type IP address, MAC address, Type and Name for the Agilent 1200 into DHCP
- Instruments tab in the node properties should be "OK? Yes" for the Agilent 1200.
- Create a system and instrument method in Empower.
- Run as usual.