How to update the Agilent PreConfiguration utility when there is a change in Agilent GC instrument hardware - WKB85496

Article number: 85496

OBJECTIVE or GOAL

Update the Agilent PreConfiguration Utility when there is a change in Agilent GC instrument hardware.

ENVIRONMENT

  • Agilent Instrument Control Framework (all versions)
  • Agilent 7890 GC
  • Agilent 7697A Headspace
  • Agilent 6890 GC
  • Agilent 6850 GC
  • Empower Enterprise

PROCEDURE

  1. Bring the system configured to Empower offline.
  2. Update the Pre-Configuration utility by selecting the "Get GC Configuration" icon for GC.
  3. If G1888 is configured, update the PreConfiguration utility by selecting the "Upload from Instrument" button on the Configure tab.
  4. If HS7697 is configured, update the PreConfiguration utility by selecting the "Upload Config from Instrument" button on the Configure tab.
  5. Bring the system configured to Empower online.
  6. Ensure that the system shows YES in Empower node properties on the Instruments tab.

ADDITIONAL INFORMATION

Note: If the system does not show YES in Empower node properties after updating the PreConfiguration utility, restart the LAC/E.