6890 controlled by Empower 3 will not communicate after relocation - WKB7425
Article number: 7425
SYMPTOMS
- After moving the system to a new location, the system will not communicate with the PC
 - The Avocent card can be pinged, but the system cannot
 
ENVIRONMENT
- Agilent 6890 Gas Chromatographer
 - Empower 3
 - Avocent hub
 
CAUSE
System configuration using the wrong port of the Avocent card
FIX or WORKAROUND
- Delete system.
 - Set all ports to "none".
 - Restart the GC.
 - Restart the PC.
 - Set the correct port.
 - Recreate the system.
 
ADDITIONAL INFORMATION
id7425, 6890A/+GC, 6890N GC, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP
 