What is the current support status of ACQUITY UPLC on Sciex Analyst? - WKB74298

Article number: 74298

ENVIRONMENT

  • ACQUITY UPLC
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC M-Class
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC I-Class PLUS
  • ACQUITY Premier
  • nanoACQUITY UPLC
  • Sciex Analyst

ANSWER

Analytical Scale ACQUITY UPLC

Waters has traditionally tested ACQUITY platforms on Analyst software. The Waters ICS is installed from the Waters driver pack media, which automatically installs a Sciex product called AAO, or Analyst Access Object. This service runs in the background and allows the ACQUITY ICS to communicate with Analyst software .

Currently, Sciex is shipping Analyst 1.7.X for tandem quadrupole instruments and Analyst 1.8.XTF for time of flight instruments. These versions will likely be installed on Windows 10, though Windows 7 support is available. Waters currently supports the following versions.

  • Analyst acquisition control (AAO, or Analyst Access Object) will be supported by the SCIEX CDS Support Layer version 2.0.0, which supports Driver Pack 2023 Release 1 for Analytical UPLC Systems with version 1.7.3 and later
  • Waters DP2022R1 was tested with and provides support for Analyst 1.7.2, 1.7.3 and 1.8.1 TF
  • Waters DP2020R1 was tested with and provides support for Analyst 1.7.2
  • Waters DP2019R1 was tested with and provides support for Analyst 1.7.0 and 1.7.1 on Windows 7 and 10
  • For previous releases of Analyst software, consult the appropriate Waters driver pack release notes for support details on www.waters.com

Multi Dimensional(2D) ACQUITY is now supported under Analyst control using only DP2022R1. The identification of pumps will take place in ACQUITY Console and then appropriate gradients for each pump can be set as normal in the instrument method. *For BSM/QSM configurations only. Two like pumps (BSM/BSM) not supported with  Analyst 1.7.2

ACQUITY UPLC M-Class, nanoACQUITY and Microscale ACQUITY UPLC

Analyst acquisition control (AAO, or Analyst Access Object) is supported by the SCIEX CDS Support Layer version 2.0.0, which supports Driver Pack 2022 Release 1 for Microscale Systems with version

  • Analyst 1.7.2
  • Analyst 1.7.3
  • Analyst 1.8.1 TF

DP2022R1 provides support for M-Class on Analyst in Windows10 with the following CDS versions.

  • Analyst 1.7.2
  • Analyst 1.7.3
  • Analyst 1.8.1 TF

Windows 7 - M-Class Driver Pack 1.55 for Third Party Support

Waters does currently test our M-Class driver pack for use with Sciex Analyst software. Per the ACQUITY UPLC MClass Driver Pack Version 1.55 for Thermo Scientific Xcalibur and Sciex Analyst Control Release Notes, full nanoACQUITY and M-Class support are offered to the following software versions.

  • Analyst 1.6.2 on Windows 7 32 bit
  • Analyst 1.6.2 on Windows 7 64 bit

Please follow the release notes for installation instructions.

Windows 10 - Driver Pack 2022 Release 1

Configuring in Analyst and Making Contact Closure Connections

For instructions on configuring the ACQUITY within Analyst, as well as making methods and running samples, please refer to the Waters LC Support Layer for Sciex CDS document attached to this article. The I/O cable from the Sciex MS must be connected to the ACQUITY to facilitate inject start and stop flow signals appropriately. Configure the connections as detailed below.

  • Inject start - connect the white/black and green/black wires to the inject start ports on the SM. Output signal is contact closure so polarity is not a concern.
  • Stop flow - connect the blue/black and blue wires to the stop flow ports on the pump. Output signal is contact closure so polarity is not a concern.
  • Connect the orange/black and red/black wires to each other and secure with wire connector cap or electrical tape.