Options tab is missing from the GC instrument method for ICF SL 3.6 in Empower - WKB233485
Article number: 233485
SYMPTOMS
- The options tab is missing from the GC instrument method in LAC/E as well as clients, including the Citrix client
- Options tab is missing from the new instrument method as well as the instrument method created on the previous ICF SL version
ENVIRONMENT
- Agilent Instrument Control Framework (ICF SL v3.6)
- Empower 3 Enterprise (including Citrix environment)
CAUSE
A dual-tower batch file script (Drive:\UtilityScripts\DualTower\Enable_Dual_Tower.bat) is installed on the LAC/E, client or on the Citrix server. - (Drive - ICF iso File)
FIX or WORKAROUND
- Run the Disable dual tower batch file script available in the Utility folder. (Drive:\UtilityScripts\DualTower\Disable_Dual_Tower.bat)
- CRI-4629 was created for this issue.
- For a dual-tower GC in a Citrix environment, it is recommended that you run the system directly on the acquisition client with the "Enable_Dual_Tower.bat" script installed on the LAC/E.
ADDITIONAL INFORMATION
Using ICF SL 3.6.1, the option tab in the Instrument method for an Agilent GC system is available whether or not Dual tower functionality is being used.