Troubleshooting module missing from ACQUITY or MS console (Guided Troubleshooting) - WKB97182

Article number: 97182

SYMPTOMS

One or multiple instrument modules are missing from the MS or ACQUITY console. 

ENVIRONMENT

  • ACQUITY console
  • ACQUITY QDa Mass Detector
  • ACQUITY UPLC
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC I-Class
  • Empower
  • MassLynx 4.1
  • MassLynx 4.2
  • MS console
  • SQ Detector 2
  • SYNAPT G2
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • SYNAPT XS
  • TQ Detector
  • Xevo QTof 
  • Xevo G2
  • Xevo G2-S
  • Xevo G2-XS
  • Xevo TQ MS
  • Xevo TQD
  • Xevo TQ-S
  • Xevo TQ-S micro
  • Xevo TQ-S cronos
  • Xevo TQ-XS

STRATEGY

SIMPLE SOLUTIONS
  1. Ensure that Windows user has full control of MassLynx and Waters Instruments folders - WKB24705
  2. Ensure that User Account Control Setting is set to "Never Notify" - WKB17373
  3. With MassLynx closed, restart MS electronics - WKB654
  4. Restart PC and all modules - WKB3194
  5. Ensure that the DHCP server configuration is set up correctly - WKB93827
  6. Delete module or modules from DHCP and repopulate - WKB4409
INTERMEDIATE SOLUTIONS
  1. Delete and regenerate ACQUITYSettings.xml file (and others, as needed) under MassLynx - WKB11966
  2. Delete and regenerate DHCP.xml file - WKB2631
  3. Inspect IntelliStart .xml files and replace corrupted ones - WKB11456
ADVANCED SOLUTIONS
  1. Install ICS as required for some standalone MS systems - WKB29933
  2. Load .net framework from appropriate driver pack - WKB23238
  3. Run ICSCompanionSetup.exe - WKB23234
  4. Disable or remove unsupported antivirus software - WKB31257
  5. Reinstall MassLynx - WKB11175
  6. Back up MassLynx data and restore PC to earlier point - WKB1557

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.