ACQUITY console error "c:\windows\microsoft.net\framework\v2.0.50727\config\machine.config Parser returned error 0xc00ce556" - WKB97128
Article number: 97128
SYMPTOMS
- When opening the console there is a quick message saying the console application is shutting down, followed by a .NET Framework error "c:\windows\microsoft.net\framework\v2.0.50727\config\machine.config Parser returned error 0xc00ce556"
ENVIRONMENT
- Any computer running MassLynx
CAUSE
Corrupted "machine.config" file
FIX or WORKAROUND
- Create a new "machine.config" file in the .NET Framework folder.
- Go to C:\\Windows\Microsoft.Net\Framework\v2.0.50727\Config\ directory, delete the corrupted machine.config file.
- Go back to the machine.config.default file and rename it to machine.config. If asked by your computer whether or not you really want to mess with the file extension, confirm the action.
- Move the new machine.config file to C:\\Windows\Microsoft.Net\Framework\v2.0.50727\Config\.
- Close File Explorer and restart the computer.