No instruments are displayed in the ACQUITY Console - WKB93827
Article number: 93827
SYMPTOMS
- No communication of instruments with the software.
- Instruments do not repopulate the DHCP server when removed.
ENVIRONMENT
- MassLynx
- Empower
- ACQUITY Console
CAUSE
IP configuration not set up correctly.
FIX or WORKAROUND
- Browse to C:\Program Files(x86)\Waters Instruments.
- Open the Waters DHCP Server Configuration.
- Remove the IP addresses from the list.
- Go to Server > Configuration Wizard.
- Select Local Area Connection, and then click <NEXT>.
- Select Default (192.168.0.1) if nothing that conflicts with those addresses, and then click <Next>.
- Click Finish.
- Restart the PC.
- Restart the ACQUITY UPLC system.
ADDITIONAL INFORMATION
If another network is using the 192.168.0.1 IP addresses, the alternate addresses (172.16.10.1) also work.