SYMPTOMS: Cannot turn on the collision gas
CAUSE: Empty argon cylinder
FIX: Replace the argon cylinder. Verify that the collision gas switch is normally on.
SYMPTOMS: MassLynx communication problem "Server Busy" error in pop-up window
CAUSE: Unsupported antivirus software is configured without the correct exceptions.
FIX: If possible, remove the unsupport...SYMPTOMS: MassLynx communication problem "Server Busy" error in pop-up window
CAUSE: Unsupported antivirus software is configured without the correct exceptions.
FIX: If possible, remove the unsupported security software and replace it with a supported version (see MassLynx release notes). If the switch is not allowed by local IT policies, ask the IT department to configure the following exceptions in the antivirus software.: C:/MassLynx/ C:/Program Files (x86)/Waters Instruments/ Any folders to
SYMPTOMS: After running samples successfully, a second set of samples submitted to the Queue to run overnight does not start Batches could be submitted from the MassLynx sample list or via OALogin Mas...SYMPTOMS: After running samples successfully, a second set of samples submitted to the Queue to run overnight does not start Batches could be submitted from the MassLynx sample list or via OALogin MassLynx stops acquiring at random intervals—sometimes immediately, sometimes in the middle of a sample set MassLynx sometimes fails during data acquisition The last data file may have no data in it
CAUSE: Illegal characters ( + - . ! @ # $ % ^ & * ( ) = and spaces) were used in the file names. See art
SYMPTOMS: Shouldering chromatographic peaks
CAUSE: Contaminated GC inlet
FIX: Perform GC inlet maintenance (replace the inlet liner and gold seal, trim the GC column at least 50 cm). If GC inlet maint...SYMPTOMS: Shouldering chromatographic peaks
CAUSE: Contaminated GC inlet
FIX: Perform GC inlet maintenance (replace the inlet liner and gold seal, trim the GC column at least 50 cm). If GC inlet maintenance fails to resolve issue consider replacing the GC column.
SYMPTOMS: Low sensitivity on customer samples Increase in sensitivity is seen when increasing the diff. app. voltage
CAUSE: Dirty Ion Path (Ion Block, StepWave or Ion Guide, Quads, Collision Cell)
FIX...SYMPTOMS: Low sensitivity on customer samples Increase in sensitivity is seen when increasing the diff. app. voltage
CAUSE: Dirty Ion Path (Ion Block, StepWave or Ion Guide, Quads, Collision Cell)
FIX: Clean or replace ion block and Ion Guide/StepWave assembly. If cleaning or replacing ion block and ion guide/stepwave does not increase sensitivity, quads and collision cell should be considered to be cleaned especially if the customer compound is relatively non-polar i.e steroids.
ANSWER: MassLynx 4.2 is supported on Windows 10. All currently shipping MS instruments are now supported with Windows 10 Enterprise LTSC 2021 (on Waters supplied Lenovo P520 and P720 only) with the la...ANSWER: MassLynx 4.2 is supported on Windows 10. All currently shipping MS instruments are now supported with Windows 10 Enterprise LTSC 2021 (on Waters supplied Lenovo P520 and P720 only) with the latest software release.Note: the release notes might not reflect the updated support for the newer Windows version. For more details choose your MS from this article:What is the latest version of MassLynx Software that supports my MS instrument? - WKB554 - Waters All currently shipping instruments ar
ANSWER: No. Whether or not this option should be used is compound-dependent and should be tested. Soft transmission may give improved sensitivity for certain compounds that fragment easily. However, i...ANSWER: No. Whether or not this option should be used is compound-dependent and should be tested. Soft transmission may give improved sensitivity for certain compounds that fragment easily. However, it may lead to a loss of sensitivity when used unnecessarily. Soft transmission mode can be activated per function in the MS method: