MassLynx does not start a sample queue or fails during acquisition - WKB2786

Article number: 2786

SYMPTOMS

  • After running samples successfully, a second set of samples submitted to the Queue to run overnight does not start
  • Batches could be submitted from the MassLynx sample list or via OALogin
  • MassLynx stops acquiring at random intervals—sometimes immediately, sometimes in the middle of a sample set
  • MassLynx sometimes fails during data acquisition
  • The last data file may have no data in it

ENVIRONMENT

  • MassLynx
  • Xevo TQ-S
  • QDa
  • Xevo TQD
  • Xevo TQ-S
  • Xevo TQ-S micro
    Xevo TQ Absolute

CAUSE

Illegal characters ( +  -  . ! @ # $ % ^ & * ( ) =   and spaces) were used in the file names.  
See article #2435 What characters can't be used in file names in MassLynx?

FIX or WORKAROUND

  1. Rename all MassLynx files (tune files, inlet methods, and MS method and sample names) so they include only the English alphabet, numbers, and the underscore.
  2. Do not use spaces in file names.
  3. Ensure that sample file names do not have spaces at the end of the name. This is difficult to spot in the sample list but can cause issues.
  4. Ensure that file text SampleID field does not contain illegal characters if the sample ID is used as the file name (OALogin).