No signal after refilling the nitrogen supply tanks - WKB27888
Article number: 27888
SYMPTOMS
- Low gas flow error message
- Nitrogen supply tanks low
- Nitrogen supply tanks refilled
- No ion beam after refilling tanks
- Found: Nitrogen gas flow in instrument OK
- Found: Source isolation valve is open
- Found: Source voltage readbacks are all correct
ENVIRONMENT
- SQ Detector (SQD)
CAUSE
Communication error
FIX or WORKAROUND
- Close MassLynx and the ACQUITY Console.
- Run StopAcquityProcesses.cmd.
- Reboot the MassLynx PC.
- Log in to Windows but to not start MassLynx.
- Push the reset button on the SQD and wait for the Operate light to go from RED to OFF.
- Open MassLynx.
ADDITIONAL INFORMATION
id27888, A-SQ, comms, SQD