MassLynx Tune Page crashes with "application error" when running Startup or Shutdown methods with operate or standby tasks - WKB279263

Article number: 279263

SYMPTOMS

  • MassLynx fails when running a startup method that contains an operate task or a shutdown method that contains the Standby task
  • Happens about 50% of the time 
  • Error message:
    "Application error 
    Application <instrument code> has encountered an unrecoverable problem and has had to exit."
    Unable to complete execution of C:\MassLynx\Shutdown\<startup method name>.acl or
    Unable to complete execution of C:\MassLynx\Shutdown\<shutdown method name>.acl or
    Unable to complete execution of shut down. Failed to switch mass spectrometer into standby
  • <instrument code> e.g. CloverUI error when running the Shutdown Method.
  • MassLynx SCN 1043 and SCN 1045
  • There may be a fluidics warning on the Tune page
  • Communications may not be lost, tune page can be reopened but MS remains in operate.
  • Maybe need reboot the computer to recover

ENVIRONMENT

  • MassLynx 
  • Xevo TQ Absolute
  • Xevo TQ-S Micro
  • Xevo TQ-S cronos
  • Xevo TQ XS
  • Xevo TQD
  • SQ Detector 2
  • ACQUITY QDa Mass Detector

CAUSE

Unknown – reported in CRI-6841 – under investigation but might be due to changes in tune page in SCN 1043

FIX or WORKAROUND

Fixed in SCN 1046 and subsequent MassLynx versions
 

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.