Multiple integrated peaks displayed in the chromatogram of the TargetLynx results - WKB27862
Article number: 27862
SYMPTOMS
- More than one peak is displayed as integrated in the chromatogram of the TargetLynx results.
- Peaks are completely resolved.
- Only one peak is annotated with the name of the compound.
- Additional peaks are annotated only with retention time and peak area.
ENVIRONMENT
- TargetLynx
- QuanLynx
- MassLynx
CAUSE
"Show All Peaks" is enabled in the display options for the chromatogram panel in the TargetLynx results.
FIX or WORKAROUND
- In the TargetLynx browser, go to Display > Options. (Alternatively, right-click chromatogram panel and select Display Options from the content sensitivity menu).
- In the Options dialog box, go to the Chromatogram tab.
- In the Style section, scroll to the end of the list.
- Disable "Show All Peaks".
- Click OK.