Should I always activate the soft transmission mode on the TQ-XS? - WKB120076
Article number: 120076
ENVIRONMENT
- Xevo TQ-XS
ANSWER
No. Whether or not this option should be used is compound-dependent and should be tested. Soft transmission may give improved sensitivity for certain compounds that fragment easily. However, it may lead to a loss of sensitivity when used unnecessarily.
Soft transmission mode can be activated per function in the MS method:
ADDITIONAL INFORMATION