Low sensitivity on Xevo and ACQUITY TQ Triple Quadrupole Instrument - WKB28620
Article number: 28620
SYMPTOMS
- Low sensitivity on customer samples
- Increase in sensitivity is seen when increasing the diff. app. voltage
ENVIRONMENT
- Xevo TQ-XS
- Xevo G2-XS
- Xevo TQ-S
- Xevo TQD
- Xevo TQ MS
- ACQUITY TQD system
CAUSE
Dirty Ion Path (Ion Block, StepWave/ Ion Guide)
FIX or WORKAROUND
- Clean or replace ion block and Ion Guide/StepWave assembly.
ADDITIONAL INFORMATION
- How to clean Ion block on Xevo instruments KCS article - https://support.waters.com/KB_Inst/M...i_style_source
- If cleaning or replacing ion block and ion guide/stepwave does not increase sensitivity, quads and collision cell should be considered to be cleaned especially if the customer compound is relatively non-polar i.e steroids.