Article number: 28711

SYMPTOMS

  • Chromatography is unsatisfactory with poor peak shape - broad peaks. 
  • When creating MS Method the Points Per Peak is lower than it should be - half compared to a working system. 
    For example: Peak width = 4 and Required Points Per Peak = 12.
  • Cannot achieve 15 points per peak even though the dwell time was adjusted to the lowest possible setting in the MS method.
  • Has been seen after a PM visit or Installation visit.

ENVIRONMENT

  • Xevo TQ-S
  • Xevo TQ-XS
  • MassLynx
  • TargetLynx

CAUSE

Inter-Scan Setup was set to Manual mode. 

FIX or WORKAROUND

  1. Disable Manual Inter-Scan Delay setting on the tune page (Tune page> Setup >Inter-Scan Setup) and ensure that "Enable Manual Mode" is NOT selected.

ADDITIONAL INFORMATION

During a PM, for the charging test in some instruments, the Inter-scan Setup is set to manual. If this value is not reset this issue can occur. Same problem can be encountered after an Installation visit. 