"Communication Failure" to Waters instrumentation within Empower 3 software (Guided Troubleshooting) - WKB233265
Article number: 233265
SYMPTOMS
- "Communication failure" to Waters instrumentation within Empower 3 software
ENVIRONMENT
- Empower 3
- Empower 3 Personal
- Empower 3 Workgroup
- Empower 3 Enterprise
STRATEGY
- Confirm creation of chromatographic system. WKB89013
- Reset modules in ACQUITY Console. WKB65394
- Stop ACQUITY Processes. WKB46988
- Determine if communication is lost. WKB230216
- Reboot PC and instrument. WKB65145
- Reset DHCP. WKB49671
- If DHCP does not populate, see DHCP does not populate entries (Guided Troubleshooting) - WKB233273.
- Delete INSTSRV.DAT. WKB8778
- Corrupt DHCP.xml file WKB118082
- Turn off antivirus software. WKB29566
- Inspect Ethernet switch and cables:
- Replace the network card (NIC). WKB4027
- Repair drivers. WKB81877
- Reinstall drivers. WKB222700
- Call Waters Expert Center or submit an iRequest online here.
ADDITIONAL INFORMATION
If instrument uses IEEE: Power-on the instrument, and then start the PC.
If instrument uses Ethernet: Start the PC, and then power-on the instrument.