"Communication Failure" to Waters instrumentation within Empower 3 software (Guided Troubleshooting) - WKB233265

Article number: 233265

SYMPTOMS

  • "Communication failure" to Waters instrumentation within Empower 3 software

ENVIRONMENT

  • Empower 3
  • Empower 3 Personal
  • Empower 3 Workgroup
  • Empower 3 Enterprise

STRATEGY

  1. Confirm creation of chromatographic system. WKB89013
  2. Reset modules in ACQUITY Console. WKB65394
  3. Stop ACQUITY Processes. WKB46988
  4. Determine if communication is lost. WKB230216
  5. Reboot PC and instrument. WKB65145
  6. Reset DHCP. WKB49671
  7. Delete INSTSRV.DAT. WKB8778
  8. Corrupt DHCP.xml file WKB118082
  9. Turn off antivirus software. WKB29566
  10. Inspect Ethernet switch and cables: 
  11. Replace the network card (NIC). WKB4027
  12. Repair drivers. WKB81877
  13. Reinstall drivers. WKB222700
  14. Call Waters Expert Center or submit an iRequest online here.

 

 

ADDITIONAL INFORMATION

If instrument uses IEEE: Power-on the instrument, and then start the PC.

If instrument uses Ethernet: Start the PC, and then power-on the instrument.