2489 UV/Vis Detector communication issues (Guided Troubleshooting) - WKB95305

Article number: 95305

SYMPTOMS

2489 UV/Vis Detector communication issues

ENVIRONMENT

2489 UV/Vis Detector

STRATEGY

  1. Reset detector from Console or Run Samples window - WKB8172
  2. Confirm that network adaptors are still bridged - WKB31622 
  3. Restart detector and LAC/E - WKB83316
  4. Inspect for antivirus software. If present, either turn off or ensure that there are exclusions from the real-time scanning - WKB29566
  5. Confirm instrument LAN Settings:
    1. Resolve communication issues - WKB67308
    2. Confirm LAN setting - WKB1972
  6. Inspect Ethernet switch and cables: 
    1. Change Ethernet switch port- WKB65949
    2. Inspect Ethernet cable - WKB78283
    3. Replace Ethernet cable - WKB11345
  7. Reinstall drivers - WKB19236

ADDITIONAL INFORMATION

 

id95305, 2489, 2489B, 2489SFC, A-10UV, comms, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP