Ethernet communication error "OK? No"; unable to reset - WKB19236

Article number: 19236

SYMPTOMS

  • Resetting the module in Communication failure with 2489 in Empower run samples does not resolve issue
  • 2489 is listed as "OK? No" on the Instruments tab in node properties
  • DHCP populates with the 2489 IP address and MAC address; Type and Name are blank
  • No antivirus running and the firewall is disabled
  • Restarting the system and the computer does not resolve the issue
  • Rebuilding DHCP does not resolve the issue

ENVIRONMENT

  • Empower 3
  • 2489 Dual-Wavelength Absorbance Detector
  • 2998 Photodiode Array Detector
  • Windows 2007
  • Configuration #15 LAC/E32 Acquisition Server
  • PCI BUS/LAC/E board installed

CAUSE

Incorrect BUS/LAC/E Driver start type

FIX or WORKAROUND

  1. Run C:\Empower\BuslaceDrivers64\SetBuslaceSvcStartType.bat.
  2. Reboot the LAC/E32.

ADDITIONAL INFORMATION


id19236, 2489, 2489B, 2489SFC, 2998, 2998B, 2998SFC, A-10PDA, A-10UV, comms