No communication with 2489 after a power failure - WKB31622
Article number: 31622
SYMPTOMS
- 2489 shows Status OK? No in the node properties.
- e2695 shows Status OK? Yes in the node properties.
- IEEE instruments all have communication.
- Restarting the instruments and LAC/E does not restore communication.
ENVIRONMENT
- 2487 Dual-Wavelength Absorbance Detector
- LAC/E32
- 2489 UV/Vis Detector
- e2695 Separations Module
- Alliance HPLC System
- Empower 3
CAUSE
Network bridge became un-bridged after power failure.
FIX or WORKAROUND
Recreate the instrument LAN bridge.
ADDITIONAL INFORMATION
id31622, comms
