Article number: 67308

SYMPTOMS

  • Unable to connect to instrument components in Empower.
  • Instrument modules are OK? NO. 
  • DHCP only populates with 2489 IP address.
  • Switching ports on the switch box did not resolve the issue.
  • Instruments initialize correctly with no hardware failures.

ENVIRONMENT

  • Empower
  • eAlliance HPLC System
  • e2695 Separations Module
  • 2489 UV/Vis Detector

CAUSE

Communication failure

FIX or WORKAROUND

  1. Power-off the instruments.
  2. Delete IP addresses from Waters DHCP. 
  3. Restart the LAC/E32 (or workstation) and log into Windows.
  4. Power-on the e2695 and 2489.
  5. Log into Empower and confirm that, in node properties, both say OK? YES.

ADDITIONAL INFORMATION

 