How often should I create a new MassLynx project? - WKB285985

Article number: 285985

ENVIRONMENT

  • MassLynx
  • TargetLynx

ANSWER

For good project management in MassLynx, a new project should be created after approximately 2500 to 3000 injections. When the number of .raw files (one .raw file = one injection) becomes too large, the indexing of the files in the project by the operating system can interfere with MassLynx's ability to access existing files or create new files.

Although TOF instrument data file folders tend to be larger, quadrupole instrument data file folders tend to be more abundant because several functions per injection are created. Each raw data file contains up to three files for each function (.dat, .IDX, .STS). 

Additionally, best practice is to back up data and move it off the PC at regular intervals to prevent full hard drives.

Tip: Create a new folder "MassLynx_Projects" (or appropriate name) and create new projects within that folder (C:\ or data drive D:\) for easy backup. For quads, you can create the Project Folder on a mapped network drive. 

ADDITIONAL INFORMATION

Symptoms of poor MassLynx project management include:

  • MassLynx randomly shows "Not Responding" status (spinning donut cursor)
  • Requires hard reboot of the PC
  • Sample list stops randomly in the middle of a batch 
  • Sample list does not advance to the next sample
  • No error message is given
  • Sample list is not "Paused"
  • Shutdown Method does not run
  • MassLynx sometimes does not inject all the samples in the sample list
  • Slow to open data files or display chromatograms
  • Display of ACQUITY modules status (such as "Injecting" or Priming") missing from the MS console
  • Erratic behavior that includes source and system settings defaulting themselves automatically even though parameters were saved on PC and EPC
  • Poor assay results after transferring project folder from older system to a different system platform.

Which characters cannot be used in file names in MassLynx? - WKB2435

How to create a project with MassLynx - WKB86161

  • Always create MassLynx projects using a template project (such as C:\MassLynx\Default.pro) that contains all the default MassLynx files in the Acqudb folder. 
  • Do not create projects "from New". The lack of default files may lead to problems with inlet control, depending on the system.

What Windows permissions does MassLynx require? - WKB94845

MassLynx sample list run stops randomly in the middle of a batch with no apparent error - wkb272176

MassLynx software randomly unresponsive, requires a hard reboot of the PC - WKB49926

Multiple errors in MassLynx including skipping samples and slow opening files - WKB8307

How to use BackLynx to back up MassLynx data - WKB100391

Insufficient disk space with MassLynx - WKB28719

How do you delete projects in MassLynx - wkb249785

Can Masslynx data be saved to more than one file location during the acquisition - wkb87174

How to setup Network Acquisition in MassLynx 4.1 or 4.2 - wkb234451

Setup for Remote Data Processing in a MassLynx and TargetLynx environment. - wkb122826