Clicking on IntelliStart from the MS Console brings "No Calibration for Selected Analysis" message - WKB285859
Article number: 285859
SYMPTOMS
- From MS Console, clicking on IntelliStart automatically generates "No Calibration for Selected Analysis" message
- Tune file was previous loaded and linked
- Automatically loads Uncal.cal
ENVIRONMENT
- MassLynx
- IntelliStart
CAUSE
The user is calibrating manually, so there are no valid IntelliStart resolution and calibration files for this analysis type. Even if these files are saved in the IntelliStart folder, it is not sufficient.
FIX or WORKAROUND
You must have an IntelliStart calibration file for each analysis type.