Clicking on IntelliStart from the MS Console brings "No Calibration for Selected Analysis" message - WKB285859

Article number: 285859

SYMPTOMS

  • From MS Console, clicking on IntelliStart automatically generates "No Calibration for Selected Analysis" message
  • Tune file was previous loaded and linked 
  • Automatically loads Uncal.cal

ENVIRONMENT

  • MassLynx 
  • IntelliStart

CAUSE

The user is calibrating manually, so there are no valid IntelliStart resolution and calibration files for this analysis type. Even if these files are saved in the IntelliStart folder, it is not sufficient.

FIX or WORKAROUND

You must have an IntelliStart calibration file for each analysis type.

ADDITIONAL INFORMATION

 