Insufficient disk space with MassLynx - WKB28719

Article number: 28719

SYMPTOMS

  • Low disk space
  • Runs cannot start
  • Sample list hangs

ENVIRONMENT

  • MassLynx

CAUSE

The C: drive is full because the MassLynx projects are being acquired to the C: drive.

FIX or WORKAROUND

  1. Open the Default project in MassLynx.
  2. Using Windows Explorer, move the projects from the C:\MassLynx folder to the data drive (D: or E:, depending on what is available). Be sure to cut and paste or copy and paste, and then delete from the C: drive.
  3. In MassLynx, open the project from the new drive.
  4. Acquire.