MassLynx software randomly unresponsive, requires a hard reboot of the PC - WKB49926
Article number: 49926
SYMPTOMS
- MassLynx randomly shows "Not Responding" status (spinning donut cursor)
- Requires hard reboot of the PC
- Found: number of .raw files in DATA folder is 365
- Found: more than 225,000 files in the DATA folder
ENVIRONMENT
- MassLynx 4.1 SCN 855
- Windows 7 Professional 64 Bit SP1
CAUSE
Number of files in the data folder is very high.
FIX or WORKAROUND
Create new project.
ADDITIONAL INFORMATION
Although the number of .raw files in the DATA folder was not high, due to the number of functions in the MS method (57), the number of individual files was very high. The indexing of this large number of files was likely the root cause of the issue.