PDA baseline noise (Guided Troubleshooting) - WKB97186
- Last updated
-
-
Save as PDF
-
SYMPTOMS
- Baseline noise in PDA chromatogram.
- SQT test failing for noise specification.
- Customer having issues with quantification at lower levels.
ENVIRONMENT
- ACQUITY UPLC PDA Detector
- ACQUITY UPLC PDA eλ Detector
- 2998 Photodiode Array Detector
- ACQUITY Arc PDA 2998
STRATEGY
- For ACQUITY PDA, inspect 250 psi back pressure regulator - WKB10081
- On ACQUITY PDA, minimize refractive index effects. If running at low UV wavelengths (less than 205 nm):
- Sonicate mobile phases - WKB1529
- For QSM, place all lines in a single mobile phase bottle - WKB95837
- If using a 30 cm column heater, insulate column outlet lines - WKB1035
- If possible, turn on Median Baseline Filter (MBF) - WKB95839
- Inspect flow cell
- Clean flow cell - WKB9482
- For 2998, rebuild flow cell - WKB13746
- If necessary, replace flow cell - WKB23351
- Replace lamp - WKB20304
- Inspect solvent manager for pressure fluctuations
- If necessary, replace check valve - WKB73348
- If necessary, rebuild pump WKB74137