Cyclic baseline on ACQUITY UPLC H-Class PDA system - WKB73348
Article number: 73348
SYMPTOMS
- Cyclic baseline when running samples on ACQUITY H-Class PDA system.
- Cleaning the instrument does not resolve the issue.
- Monitoring baseline with no flow does not indicate an electronic issue.
- PDA energy looks good.
- When monitoring baseline, accumulator pump trace looks like there is an issue with the check valve.
ENVIRONMENT
- ACQUITY UPLC H-Class
- ACQUITY UPLC PDA Detector
- ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)
CAUSE
QSM accumulator check valve is sticking.
FIX or WORKAROUND
Replace the accumulator check valve.
ADDITIONAL INFORMATION
QSM accumulator check valve: 700005164 (CHECK VALVE, DOUBLE BALL AND SEAT, 1PK)