SYMPTOMS

  • Cyclic baseline when running samples on ACQUITY H-Class PDA system. 
  • Cleaning the instrument does not resolve the issue.
  • Monitoring baseline with no flow does not indicate an electronic issue.
  • PDA energy looks good. 
  • When monitoring baseline, accumulator pump trace looks like there is an issue with the check valve.

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • ACQUITY UPLC PDA Detector
  • ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)

CAUSE

QSM accumulator check valve is sticking.

FIX or WORKAROUND

Replace the accumulator check valve.

ADDITIONAL INFORMATION

QSM accumulator check valve: 700005164 (CHECK VALVE, DOUBLE BALL AND SEAT, 1PK)

id73348, A-30PDA, checkvalve, UPAPDA, UPC2PDA, UPPDA, UPPDAARC, UPPDAARCB, UPPDA-E, UPPDA-L, UPPDALTC, UPPDATC, UPPPDA, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO