All SEC peaks are getting broader on ACQUITY H-Class Bio - WKB73253
Article number: 73253
SYMPTOMS
- All SEC peaks are getting broader over time or showing double peaks
- Resolution of separation is no longer good
ENVIRONMENT
- ACQUITY UPLC H-Class Bio
- ACQUITY UPLC
- Size-exclusion chromatography
CAUSE
Microbial growth on column
FIX or WORKAROUND
- Remove the contaminated solvent bottles and column.
- Clean your solvent lines, pump, and injector using phosphoric acid wash (30%). See Controlling Contamination in LC/MS Systems.
- Clean your detector according to the instructions in its operator manual (e.g., How to clean the ACQUITY PDA Detector flow cell).
- Rinse the entire system with UPLC-grade water in a clean glass borosilicate bottle.
- Make fresh solvent in clean bottles, fit a new SEC column, and set up the system with your SEC application according to the care and use manual of the column (ACQUITY UPLC Protein BEH SEC Columns and Standards).
- To avoid microbial growth, follow guidelines in Guidelines for routine use and maintenance of ultra-performance size-exclusion and ion-exchange chromatography systems and Size exclusion and ion-exchange chromatography of proteins using the ACQUITY UPLC H-Class System.
ADDITIONAL INFORMATION
id73253, UPBINARY, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO