Baseline noise is observed at low UV wavelengths - WKB95839

Article number: 95839

SYMPTOMS

  • Baseline noise observed at low UV wavelengths (under 205 nm).

ENVIRONMENT

  • ACQUITY UPLC TUV Detector
  • ACQUITY UPLC PDA Detector

CAUSE

Refractive index changes are enhanced at low UV wavelengths. 

FIX or WORKAROUND

Acquire the data using Median Baseline Filter (MBF).