Baseline steps on ACQUITY ARC PDA (UPPDAARC) - WKB20304
Article number: 20304
SYMPTOMS
- Baseline steps (see example in additional information below).
- No pressure variability corresponding to the steps.
ENVIRONMENT
- ACQUITY Arc PDA (UPPDAARC)
- 2998 Photodiode Array Detector
- ACQUITY UPLC PDA Detector
- ACQUITY UPLC PDA eλ Detector
CAUSE
Defective lamp.
FIX or WORKAROUND
Replace the lamp (700009330).
ADDITIONAL INFORMATION
id20304, 2998, 2998B, 2998SFC, A-10PDA, A-30PDA, UPAPDA, UPAQSM, UPARBQSM, UPARBSMFTN, UPARCH30, UPARHQSM, UPARPPDA, UPARQSM, UPARSMFTN, UPC2PDA, UPPDA, UPPDAARC, UPPDAARCB, UPPDAARCB, UPPDA-E, UPPDA-L, UPPDALTC, UPPDATC, UPPPDA