Leak detected in ACQUITY UPLC Sample Manager - WKB203041

Article number: 203041

SYMPTOMS

  • Leak detected
  • Leak seen at port 5 of the injection valve

ENVIRONMENT

  • ACQUITY UPLC 
  • ACQUITY UPLC I-Class

CAUSE

The pump outlet tubing is worn down.

FIX or WORKAROUND

Replace the pump outlet tubing.

ADDITIONAL INFORMATION

  • ACQUITY UPLC I-Class - Tubing, BSM Outlet to SM (# 430002901) 
  • ACQUITY UPLC - System Outlet Tube Assembly (# 430001486)