Article number: 202995

SYMPTOMS

  • Leak where manifold tubing meets I2V

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class PLUS

CAUSE

Damaged or worn fitting on I2V

FIX or WORKAROUND

  1. Tried replacing manifold tubing but leak remained. PN: 430002400
  2. Replaced I2V and leak stopped. PN: 700005162

ADDITIONAL INFORMATION