Fluctuating Baseline on UV/VIS Detectors - WKB97190

Article number: 97190

SYMPTOMS

Wavy/Noisy baseline. 

ENVIRONMENT

  • 2487 UV/VIS detector
  • 2489 UV/VIS detector
  • ACQUITY Arc 2489 detector
  • 2996 PDA detector
  • 2998 PDA detector
  • ACQUITY Arc 2998 detector
  • ACQUITY TUV detector
  • ACQUITY PDA detector
  • ACQUITY UPC2 PDA Detector

CAUSE

Detector lamp failing

FIX or WORKAROUND

Replace the lamp.

  • Lamp for 2487 detector and ACQUITY TUV detector (through serial number K05UPT699N) - WAS081142
  • Lamp for 2489 Platinum detector, 2998 Platinum and Blue detectors, ACQUITY Arc 2998 detector, ACQUITY TUV detector (serial number: K05UPT700N and later) and ACQUITY PDA eLambda detector - 201000281 
  • Lamp 2489 Blue detector and ACQUITY Arc 2489 detector - 700009330
  • Lamp for 2996 PDA / ACQUITY 2996 detector - WAT052586
  • Lamp for ACQUITY TUV detector (serial number: A10TUV001N and later) and ACQUITY PDA eLambda detector (Thermally managed) and ACQUITY UPC2 PDA Detector Part Number 700005269

ADDITIONAL INFORMATION

Example of a wavy/noisy baseline.

clipboard_ebd33164c0e74925e35d5268389c23857.png

Not able to find a solution? Click here to request help.