Resetting the SM-FTN turns off the CHA set temperature - WKB97198

Article number: 97198

SYMPTOMS

  • Resetting the SM-FTN turns off the CHA set temperature
  • No effect on sample temperature

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • Sample Manager - FTN
  • Firmware 1.71

CAUSE

SM-FTN Firmware 1.71 bug

FIX or WORKAROUND

Use a version other than SM-FTN ICS 1.71 (1.65 or other).

However, SM-FTN Plus cannot be supported because FW 1.71 or higher is required.

[Reference] DP2019R2 (ICS1.74 / FW1.71), DP2019R1 (ICS1.71 / FW1.71), DP2018R1 (ICS1.71 / FW1.71), DP2017R1 (ICS1.69 / FW1.65).

ADDITIONAL INFORMATION