2998 PDA detector - excessive baseline noise - WKB13746

Article number: 13746

SYMPTOMS

  • Excessive, irregular baseline noise

ENVIRONMENT

  • 2998 PDA detector

CAUSE

Poor flow cell condition

FIX or WORKAROUND

  1. Rebuild the flow cell using the appropriate rebuild kit. 
  2. Replace flow cell if rebuilding the flow cell does not resolve the baseline issue.
  3. Run Noise / Drift test to confirm functionality 

 

ADDITIONAL INFORMATION

You can apply the resolution described here to a failure of the UV detector SQT noise and drift test.

Platinum Detector Related Parts:
Flow Cell, Analytical TaperSlit, 10 mm, 8.4 µL (# 205000399) 
Flow Cell, Autopurification, 0.5 mm, 12.3 µL (# 205000402) 
Flow Cell, Microbore TaperSlit, 8 mm, 2.7 µL (# 205000400) 
Flow Cell, Semi Prep TaperSlit, 3 mm, 16.3 µL (# 205000401) 
 Accessories:
Flow Cell Rebuild Kit, Analytical, 10 mm (# 700003798) 
Flow Cell Rebuild Kit, Autopurification, 0.5 mm (# 700005102) 
Flow Cell Rebuild Kit, Semi Prep TaperSlit, 3 mm (# 700005101) 

Blue Detector Related Parts:
Flow Cell, Analytical TaperSlit, 10 mm, 8.4 µL (# 205001023) 
Flow Cell, Autopurification, 0.5 mm, 12.3 µL (# 205001026) 
Flow Cell, Microbore TaperSlit, 8 mm, 2.7 µL (# 205001024) 
Flow Cell, Semi Prep TaperSlit, 3 mm, 16.3 µL (# 205001025) 
 Accessories:
Flow Cell Rebuild Kit, Analytical, 10 mm (# 700003798) 
Flow Cell Rebuild Kit, Autopurification, 0.5 mm (# 700005102) 
Flow Cell Rebuild Kit, Semi Prep TaperSlit, 3 mm (# 700005101) 

Arc Detector Related Parts:
Flow Cell, Low Dispersion, 10 mm, 8.4 µL (# 205001552) 
Flow Cell, Inert, Arc Bio (# 205001043) 
 Accessories:
Flow Cell Rebuild Kit, Analytical, 10 mm (# 700003798) 

id13746, 2998, 2998B, 2998SFC, A-10PDA, baseline ripple