2998 PDA detector - excessive baseline noise - WKB13746
SYMPTOMS
- Excessive, irregular baseline noise
ENVIRONMENT
- 2998 PDA detector
CAUSE
Poor flow cell condition
FIX or WORKAROUND
- Rebuild the flow cell using the appropriate rebuild kit.
- Replace flow cell if rebuilding the flow cell does not resolve the baseline issue.
- Run Noise / Drift test to confirm functionality
ADDITIONAL INFORMATION
You can apply the resolution described here to a failure of the UV detector SQT noise and drift test.
Platinum Detector Related Parts:
Flow Cell, Analytical TaperSlit, 10 mm, 8.4 µL (# 205000399)
Flow Cell, Autopurification, 0.5 mm, 12.3 µL (# 205000402)
Flow Cell, Microbore TaperSlit, 8 mm, 2.7 µL (# 205000400)
Flow Cell, Semi Prep TaperSlit, 3 mm, 16.3 µL (# 205000401)
Accessories:
Flow Cell Rebuild Kit, Analytical, 10 mm (# 700003798)
Flow Cell Rebuild Kit, Autopurification, 0.5 mm (# 700005102)
Flow Cell Rebuild Kit, Semi Prep TaperSlit, 3 mm (# 700005101)
Blue Detector Related Parts:
Flow Cell, Analytical TaperSlit, 10 mm, 8.4 µL (# 205001023)
Flow Cell, Autopurification, 0.5 mm, 12.3 µL (# 205001026)
Flow Cell, Microbore TaperSlit, 8 mm, 2.7 µL (# 205001024)
Flow Cell, Semi Prep TaperSlit, 3 mm, 16.3 µL (# 205001025)
Accessories:
Flow Cell Rebuild Kit, Analytical, 10 mm (# 700003798)
Flow Cell Rebuild Kit, Autopurification, 0.5 mm (# 700005102)
Flow Cell Rebuild Kit, Semi Prep TaperSlit, 3 mm (# 700005101)
Arc Detector Related Parts:
Flow Cell, Low Dispersion, 10 mm, 8.4 µL (# 205001552)
Flow Cell, Inert, Arc Bio (# 205001043)
Accessories:
Flow Cell Rebuild Kit, Analytical, 10 mm (# 700003798)
id13746, 2998, 2998B, 2998SFC, A-10PDA, baseline ripple