Troubleshooting poor mass accuracy on TOF instruments with IntelliStart (Guided Troubleshooting) - WKB96600

Article number: 96600

SYMPTOMS

Symptoms are not limited to but may include:

  • Mass accuracy on standards and/or customer's samples is consistently poor
  • Installation or qualification mass accuracy test fails
  • LCMS system check fails due to poor mass accuracy
  • Excessive chemical noise interferes with some peaks of interest

ENVIRONMENT

  • Xevo QTof
  • Xevo G2
  • Xevo G2-S
  • Xevo G2-XS
  • SYNAPT G2
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • SYNAPT XS

STRATEGY

Software Solutions:

  1. If applicable, ensure that MassLynx data is processed using Automatic Peak Detection and correct LockMass settings - WKB10191
  2. Ensure that LockSpray signal is present at an appropriate intensity. If not, purge fluidics  - WKB16318
  3. Check resolution and tune as needed. Then recalibrate and run LockSpray setup - WKB17015
  4. Optimize LockSpray using scan time and frequency - WKB21899

Sample and Solvent Solution:

Use correct combination of Leu Enk and Glu Fib concentrations in LockSpray sample - WKB7813

 

Hardware Solution:

Resolve high lab temperature variation and then recalibrate - WKB13352

ADDITIONAL INFORMATION

 

id96600, MALDIG2SI, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, XEVOG2QTOF, XEVOG2TOF, XEVOQTOF

Not able to find a solution? Click here to request help.