Troubleshooting poor mass accuracy on TOF instruments with IntelliStart (Guided Troubleshooting) - WKB96600
Article number: 96600
SYMPTOMS
Symptoms are not limited to but may include:
- Mass accuracy on standards and/or customer's samples is consistently poor
- Installation or qualification mass accuracy test fails
- LCMS system check fails due to poor mass accuracy
- Excessive chemical noise interferes with some peaks of interest
ENVIRONMENT
- Xevo QTof
- Xevo G2
- Xevo G2-S
- Xevo G2-XS
- SYNAPT G2
- SYNAPT G2-S
- SYNAPT G2-Si
- SYNAPT XS
STRATEGY
Software Solutions:
- If applicable, ensure that MassLynx data is processed using Automatic Peak Detection and correct LockMass settings - WKB10191
- Ensure that LockSpray signal is present at an appropriate intensity. If not, purge fluidics - WKB16318
- Check resolution and tune as needed. Then recalibrate and run LockSpray setup - WKB17015
- Optimize LockSpray using scan time and frequency - WKB21899
Sample and Solvent Solution:
Use correct combination of Leu Enk and Glu Fib concentrations in LockSpray sample - WKB7813
Hardware Solution:
Resolve high lab temperature variation and then recalibrate - WKB13352
ADDITIONAL INFORMATION