Poor mass accuracy on a Xevo G2-XS in UNIFI - WKB21899
Article number: 21899
SYMPTOMS
- The mass accuracy is out of specification
- The analysis method has a LockSpray profile
- in the investigate tab, within the Reference Function, the lock mass is absent (possibly one out of two in a dual spray lock mass profile)
Method/LockSpray profile processes with m/z 278.0113 and m/z 556.5766 in positive ion mode
In the reference function, m/z 556.2766 is present, but m/z 278.0113 is missing
In the review tab, components are missing or show a high mass error (if target by mass filter threshold is increased to 100 ppm in the analysis method)
ENVIRONMENT
- Xevo G2-XS
- Xevo G2-S
- UNIFI 1.9.4
CAUSE
Lock mass absent. LockSpray Profile needs optimization
FIX or WORKAROUND
- Run the LockSpray setup for the profile in question.
- Open the Analysis Method and, on the MS LockSpray tab, import the LockSpray profile that was just updated.
- Do a test acquisition and verify all selected lock masses are present under method conditions
- If required, optimize flow rate, Capillary Voltage or ColIision energy in the LockSpray profile. Then run LockspraySetup and import settings to achieve a lock mass signal of 2e4 to 3e5 per second during the acquisition.
- Save or update the method and acquire.
ADDITIONAL INFORMATION