Pressure is too high after cleaning sample cone - WKB219635
Article number: 219635
SYMPTOMS
- Pressure too high
- Turbo does not reach 100%
- Vacuum OK when isolation valve is closed
ENVIRONMENT
- SQD
- TQD
CAUSE
Leak around the sample cone
FIX or WORKAROUND
- Refit the sample cone and the O-ring.
- If necessary, replace the sample cone O-ring (700001194).
ADDITIONAL INFORMATION
