Pressure is too high after cleaning sample cone - WKB219635

Article number: 219635

SYMPTOMS

  • Pressure too high
  • Turbo does not reach 100%
  • Vacuum OK when isolation valve is closed

ENVIRONMENT

  • SQD 
  • TQD

CAUSE

Leak around the sample cone

FIX or WORKAROUND

  1. Refit the sample cone and the O-ring.
  2. If necessary, replace the sample cone O-ring (700001194).

ADDITIONAL INFORMATION

 

id219635, A-SQ, SQD