Poor mass accuracy on Xevo G2 Tof data - WKB10191

Article number: 10191

SYMPTOMS

  • 10 to 50 ppm mass accuracy on nicely resolved data that has the lockmass applied after acquisition

ENVIRONMENT

  • MassLynx 4.1
  • Xevo G2 Qtof

CAUSE

Data is not centered properly. For G2 data, you must use automatic peak processing. 

FIX or WORKAROUND

  1. In the Spectrum window, select Process > Automatic Peak Detection.

Automatic Peak Detection 1.PNG

  1. Ensure that settings match the following, and set the correct Lock Mass species.

Automatic Peak Detection 2.PNG

ADDITIONAL INFORMATION

If you are centering data on a processing PC, ensure that the settings are as described in this article: Combine Spectrum results are different in MassLynx Desktop and Acquisition SCNs

id10191, MLYNX, MLYNXV41, SUPMM, XEVOG2QTOF, XEVOG2TOF