Mass accuracy measurement fails on LCMS data with a Xevo G2-S - WKB13352
Article number: 13352
SYMPTOMS
- Mass error measurement on m/z 152 (acetaminophen) during system precision test on "6 mix" solution fails at more than 50 ppm
ENVIRONMENT
- Xevo G2-S QTof
- Xevo G2-S Tof
CAUSE
Very high laboratory temperature owing to issue with room temperature control
FIX or WORKAROUND
- Ensure that the lab temperature meets requirements (<28 °C, short-term (1.5 h) variation < +/- 2 °C).
- Recalibrate the system.
- Rerun the precision test.
ADDITIONAL INFORMATION