Mass accuracy measurement fails on LCMS data with a Xevo G2-S - WKB13352

Article number: 13352

SYMPTOMS

  • Mass error measurement on m/z 152 (acetaminophen) during system precision test on "6 mix" solution fails at more than 50 ppm

ENVIRONMENT

  • Xevo G2-S QTof
  • Xevo G2-S Tof

CAUSE

Very high laboratory temperature owing to issue with room temperature control

FIX or WORKAROUND

  1. Ensure that the lab temperature meets requirements (<28 °C, short-term (1.5 h) variation < +/- 2 °C).
  2. Recalibrate the system.
  3. Rerun the precision test.

ADDITIONAL INFORMATION

id13352, XEVOG2SQTF, XEVOG2STOF