No MS signal using the SQD2, but the PDA is fine - WKB9660
Article number: 9660
SYMPTOMS
- Peaks appear as expected in the PDA trace
- No peaks, only noise in the MS trace
- No error messages
- No leaks
- Tune page readbacks are okay
- The ACQUITY H-Class system and the PC were rebooted
ENVIRONMENT
- SQ Detector 2
- ACQUITY H-Class system
- ACQUITY UPLC I-Class
- MassLynx 4.1
CAUSE
The EPC needs a reboot.
FIX or WORKAROUND
- Exit the console and all MassLynx windows.
- Reboot the EPC.
ADDITIONAL INFORMATION
Instructions for rebooting EPC can be found in WKB7776
id9660, A-SQ, MLYNX, MLYNXV41, SQD2, SUPMM, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO