How to optimize for either ESI or APCI using ESCI with IntelliStart - WKB9654
Article number: 9654
OBJECTIVE or GOAL
Use IntelliStart to determine whether ESI or APCI ionization is better for a particular sample.
ENVIRONMENT
- Xevo
- Xevo TQD
PROCEDURE
- Open the MS tune page.
- Select ESCI as the ionization mode.
- Open the console and select IntelliStart.
- IntelliStart will show the option to develop a method for ESI+, ESI-, APCI+, and APCI-, as long as ESCI is chosen on the MS tune page.
- Check the boxes to either develop an ESI mode method or an APCI mode method. They cannot be run at the same time.
- Run one mode as normal then run again in the other mode to generate a comparison.
- Select the MS ionization method with the strongest response to run your sample.
- As long as the ESCI mode is selected on the tune page, either ES +/- or APCI +/- methods can be run as a MS Method in the sample list.
ADDITIONAL INFORMATION
- Corona pin must be fitted to run APCI.
- If you select all four modes to run, no error will be generated but only an ESI method will be developed.
- APCI samples run with the ESi probe will optimize best if the probe is halfway between the cone and the corona pin.
- If APCI mode looks good while running ESCI, using an APCI probe will increase the sensitivity due to additional drying capabilities.
id9654, Electrospray, XEVOTQD, XEVOTQDIVD