How to know the right time to put your LCT PREMIER or Q-Tof PREMIER into Operate mode - WKB965

Article number: 965

OBJECTIVE or GOAL

Know when to put an LCT or Q-Tof PREMIER into Operate mode.

ENVIRONMENT

  • LCT PREMIER
  • Q-Tof PREMIER
  • MassLynx 4.1

PROCEDURE

  1. Start the pump-down via Tune page > Vacuum > Pump.
  2. Wait at least one hour.
  3. Ensure that the Tof vacuum is less than or equal to 1.2 e-6 mbar.
  4. Put the instrument into Operate mode.
  5. Condition the MCPs.

ADDITIONAL INFORMATION