IntelliStart LC/MS system check fails - WKB96480
Article number: 96480
SYMPTOMS
- LC/MS System Check fails all criteria with zero values
ENVIRONMENT
- IntelliStart
CAUSE
Incorrect LC or MS methods, functions, or channels are being used in the IntelliStart LC/MS System Check system parameters.
FIX or WORKAROUND
- Locate which tune page (.ipr) LC methods and MS method you want to use.
- Using Windows Explorer, copy those methods into C:\Masslynx\SystemQC.pro\Acqudb.
- Open the MS Console and select IntelliStart from the mass spectrometer menu.
- Select the box next to LC/MS System Check and click the "Start" balloon on the right-hand side of the console.
- From the LC/MS System Check tab, locate the method section in the bottom-left corner of the screen. For each method (Tune, LC, and MS), browse to the correct method located in the C:\Masslynx\SystemQC.pro\Acqudb folder. Additionally, choose the correct MRM parent/daughter masses and save the method.
ADDITIONAL INFORMATION
id96480, MLYNX, MLYNXV41, SUPMM