Calibration is not passing on a Xevo G2-XS instrument - WKB96452
Article number: 96452
SYMPTOMS
Calibration is not passing and lower-end masses are off.
ENVIRONMENT
- Xevo G2-XS
- MassLynx
CAUSE
Signal strength is too high in intensity.
FIX or WORKAROUND
De-tune the instrument to a low e-5 intensity level of the calibrant.
ADDITIONAL INFORMATION
You can use capillary voltage to de-tune the signal.
id96452, XEVOG2XSTF