No beam (no peaks, no signal) on TOF mass spectrometers (Guided Troubleshooting) - WKB96390

Article number: 96390

SYMPTOMS

  • No beam in some or all modes (MS, MSMS, positive, negative, others)

ENVIRONMENT

  • Xevo QTof
  • Xevo G2 QTof
  • Xevo G2-S
  • Xevo G2-XS
  • Synapt HDMS
  • SYNAPT G2-S 
  • SYNAPT G2-Si
  • Vion IMS QTof
  • ACQUITY RDa Detector
  • BioAccord
  • Q-Tof Premier

STRATEGY

  1. Increase capillary voltage on sample probe to 5 kV. If some peaks appear in low mass region, it indicates that mass spectrometer is working. Check whether fluid is supplied to the probe.
    1. Determine if the fluidics lines are out of solvent in the wash solvent reservoir. (WKB1269)
    2. Purge the sample valve. (WKB24596)
  2. Check the various modes (such as MS, MSMS, positive, negative).
  3. No beam in MSMS mode only. (WKB3503)

ADDITIONAL INFORMATION

 

id96390,

Not able to find a solution? Click here to request help.