Skip to main content
Waters

No beam when infusing to the sample probe - WKB1269

Article number: 1269

SYMPTOMS

  • No beam when infusing to the sample probe using the on-board fluidics
  • There is a beam when you have LC flow to the sample probe

ENVIRONMENT

  • Xevo G2 
  • Xevo G2-S
  • Xevo G2 XS
  • Xevo TQ MS
  • Xevo TQ-S
  • Xevo TQD
  • SQD
  • SQD2
  • Xevo TQ-S micro
  • Xevo TQ-XS
  • SYNAPT G2
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • Xevo QTof

CAUSE

The fluidics lines are out of the solvent in the wash solvent reservoir.

FIX or WORKAROUND

  1. There is an orange line that goes from the sample fluidics valve (the valve above the fluidics A, B, and C bottles) above the MS to a wash solvent reservoir.

wash_line.png

The orange lines sometimes pop up and no longer sit in the wash solvent. 

  1. Push the orange line well into the wash solvent bottle, so that it extends well below the solvent line. 
  2. If the line popping up is a frequent issue, you can take a pre-slit LC vial cap and run it up the line to the inside of the solvent bottle cap. This should help hold the line in place. 
  3. Now that the line is in the wash solvent, purge four times to remove air from the lines and valve.
  4. Try infusing. You should see your sample after a few minutes of infusing at 10 uL/min.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.