No beam or solution seen when pumping from fluidics bottle C on the Xevo G2-XS QTof - WKB24596
Article number: 24596
SYMPTOMS
- No beam when pumping from fluidics bottle C to the sample probe
- No solvent seen when disconnecting the line to the probe while pumping from C
- Purged many times
- A beam is seen on the reference probe using the fluidics
- No beam is seen when pumping from A, B, or C using the sample fluidics
- Clogged fluidics valve
ENVIRONMENT
- SYNAPT G2
- SYNAPT G2 MS/HDMS
- SYNAPT G2-S MS/HDMS
- SYNAPT G2-Si MS/HDMS
- SYNAPT G2-S
- SYNAPT G2-Si
- Vion IMS QTof
- Xevo QTof
- Xevo Tof
- Xevo G2 QTof
- Xevo G2 Tof
- Xevo G2-S
- Xevo G2-S QTof
- Xevo G2-S Tof
- Xevo G2-XS
- Xevo G2-XS QTof
- Xevo G2-XS Tof
CAUSE
The sample or diverter valve is clogged.
FIX or WORKAROUND
- Label the tubing going into the sample (middle) valve.
- Disconnect the tubing.
- Unscrew the three screws that secure the front face of the valve (stator).
- Flush the rotor (brown disk with curved groves that sits behind the stator) using a spray bottle and methanol.
- Take an Eppendorf syringe and a 1-mL tip and suck up a mL of methanol or methanol/water. Push the solution through all seven ports. Make sure the solution sprays out of the back side in a nice stream. If it sputters, repeat spraying through that port.
- Rebuild the valve, putting the stator back on, and then give each screw one turn before going back to screwing the first screw in again. It must be tightened evenly.
- Connect the tubing.
- Purge three times with IPA to get rid of air bubbles.
- Purge two times with the sample solution in C.
- Check for a beam
ADDITIONAL INFORMATION
The same can be done for the diverter (right) valve or the lockspray (left) valve.