Cant get commuication with MS and LC: The MS embedded computer (EPC) is caught in an endless loop waiting for Waters DHCP server to respond when rebooted - WKB24612
Article number: 24612
SYMPTOMS
- No communications with mass spectrometer or ACQUITY UPLC system
- No components of the system show up on the ACQUITY MS Console
- Swapping network cables does not resolve the issue
- Lights for each Ethernet port on the back of the sample manager show activity
- Rearranging the cables in the ports on the network hub does not resolve the issue
ENVIRONMENT
- Xevo TQ-S
- ACQUITY UPLC
- ACQUITY UPLC H-class
- MassLynx 4.1
- MassLynx 4.2
- Xevo QTof
- Xevo G2 QTof
- Xevo G2 Tof
- Xevo G2-S
- Xevo G2-S QTof
- Xevo G2-S Tof
- Xevo G2-XS
- Xevo G2-XS QTof
- Xevo G2-XS Tof
CAUSE
Corruption in the Waters DHCP Server Configuration
FIX or WORKAROUND
- Browse to C:\Program Files(x86)\Waters Instruments.
- Open the Waters DHCP Server Configuration.
- Remove the IP addresses from the list.
- Go to Server > Configuration Wizard.
- Select Local Area Connection <NEXT>
- Select Default (192.168.0.1) <Next>
- Hit Finish
- Reboot the PC.
- Restart the ACQUITY UPLC system.
- Restart the Xevo TQ-S electronics using the toggle switches on the back.
ADDITIONAL INFORMATION